The use of a multi-segment model of the spatial characteristics of the transformation to improve the accuracy of measurements hot-wire gauges liquid flow rate

Abstract

The use of a multi-segment model of the spatial characteristics of the transformation to improve the accuracy of measurements hot-wire gauges liquid flow rate

Klevtsov S.I., Udod E.V.

Incoming article date: 20.09.2014

To achieve high accuracy of measurement of liquid flow rate using the hot-wire gauges are encouraged to use multi-segment method of spatial characteristics of the transformation. The method is based on the nonlinear three-dimensional elements. Model the spatial characteristics of the transformation is maximally adapted to the peculiarities of the conversion function of the primary transmitter. Model repeats its spatial form with the non-linearity, zero drift, the influence of external factors, including temperature. This provides a low error rate calculation of flow. Study of the effectiveness of the method was based on experimental data obtained during the calibration at different temperatures. Maximum reduced measurement error does not exceed ~ 0.45% at an arbitrary temperature, recorded during measurement.

Keywords: thermoanemometric meter spatial multi segment conversion characteristic, the error rate of fluid flow, the approximation method of least squares